![]() In the second patient group, a total of 27 vertebrae were analysed in 10 patients showing weak Pearson correlation and Bland-Altman analysis indicated no equivalence. Bland-Altman analysis indicated consistency and equivalence with a bias of 3.9 and 95% limits of agreement from -27.17 to 34.97 when comparing all pre- and intraoperative HU values of L1-5. In the first patient group, a total of 250 vertebrae were analysed in 74 patients showing a strong Pearson correlation of >.94 between pre- and intraoperative HU values. In a subgroup analysis of patient group one, radiation by means of CT Dose Index (CTDI) was compared between Canon and AIRO CT scanners. In patient group two, Canon CT scanners were used preoperatively and the O-arm Cone Beam CT (CBCT) scanner was used intraoperatively. In patient group one, Canon CT scanners were used preoperatively and the AIRO CT scanner was used intraoperatively. HU values of lumbar vertebrae were measured and compared between preoperative and intraoperative CT scans in patients undergoing lumbar interbody fusion. To evaluate the accuracy of intraoperatively measured computed tomography (CT) Hounsfield unit (HU) values by comparison with preoperative CT HU values and to compare the radiation exposure between preoperative and intraoperative CT scans.